Determination of tetrandrine in human plasma by high-performance liquid chromatography

Tan, G.S.; Dai, Z.Y.; Xu, P.S.

Hunan Yi Ke da Xue Xue Bao 25(4): 409-410

2000


ISSN/ISBN: 1000-5625
PMID: 12206020
Document Number: 515309
The paper reports the determination of tetrandrine by high performance liquid chromatography (HPLC). A C18 column was used and the mobile phase was methanol-water(80:20) containing 0.03% triethylamine, detective wave was at 282 nm, the internal standard was diazepam, the sample was extracted with ether. The linear range was from 0.289 microgram.ml-1 to 4.618 micrograms.ml-1, the average recovery rate was 95.8%. The coefficients of variation of within-day and day-to-day were less than 5%. This method is simple, rapid and accurate. It can be used in biomedical sample analysis of tetrandrine.

Document emailed within 1 workday
Secure & encrypted payments