Simultaneous determination of catecholamine levels in plasma by high performance liquid chromatography

Peng, G.H.; Tan, Y.; Zeng, L.M.

Hunan Yi Ke da Xue Xue Bao 26(5): 485-487

2001


ISSN/ISBN: 1000-5625
PMID: 12536512
Document Number: 535056
To use a high performance liquid chromatography with electrochemical detection to simultaneously determine catecholamine(CA) levels in plasma and to diagnose chromaffin cell tumors and neuroblastoma. The plasma samples after flowing extraction by ion-moderated partition were determined with electrochemical method to detect CA levels in plasma; the lowest detective limit, the precision, recovery, sensitivity of CA levels were tested and a reference range was established based on the respective data of 18 healthy persons. The recovery rates of epinephrine and norepinephrine were 86.2% and 90.5%. The method of testing epinephrine was linear at the range of 0.2 to 10.6 nmol.L-1 with 0.05 nmol.L-1 of the lowest detective limit. The intra- and inter-assay coefficients of variation were lower than 10.7% and 11.1%, respectively. The method is simple, accurate, sensible in the clinical diagnosis of CA levels.