The determination of creatinine in blood and plasma of small animals using the dry chemical system Reflotron
Kridde, G.; Hirschberger, J.
Tierarztliche Praxis 19(4): 447-449
1991
ISSN/ISBN: 0303-6286 PMID: 1948995 Document Number: 381296
The reagent carrier Reflotron-Creatinine was tested for its suitability for whole blood and plasma samples of dogs and cats. It showed a very good correlation between the results of Reflotron-Creatinine and the reference method. Using plasma the results of both methods were extremely close. Using whole blood with a hematocrit of up to 50% good results could be achieved. If the hematocrit exceeded 50% large deviations could result.