Determination of retinol in human blood plasma using high pressure liquid chromatography

Sergeev, A.V.; Shevchenko, V.E.; Kaganovich, M.V.

Voprosy Meditsinskoi Khimii 29(5): 54-58

1983


ISSN/ISBN: 0042-8809
PMID: 6139916
Document Number: 219845
A quantitative procedure for estimation of retinol in human blood plasma using high pressure liquid chromatography involved separation of the retinoids on an reversed phase analytical column. C19-retinoid aldehyde was used as an internal standard. The procedure enabled to detect up to 0.015 microgram of retinol/ml of blood plasma and may be used for routine clinical analyses.