Assay of underivatized intrazepam and clonazepam in plasma by capillary gas chromatography applied to pharmacokinetic and bioavailability studies in humans

de Boer, A.G.; Röst-Kaiser, J.; Bracht, H.; Breimer, D.D.

Journal of Chromatography 145(1): 105-114

1978


ISSN/ISBN: 0021-9673
PMID: 621234
Document Number: 137074
The assay procedure of underivatized, intact nitrazepam and clonazepam in human plasma is described, using gas chromatography with a support-coated open tubular column (OV-17), a solid injection system and electron-capture detection. Clonazepam is used as internal standard in the assay of nitrazepam and vice versa. Linear calibration curves after a single extraction step were obtained in the concentration range 10-100 ng/ml plasma, with standard deviations less than 4.9%. The sensitivity limit of the method is about 1 ng/ml plasma for both drugs. The method was applied to pharmacokinetic and bioavailability studies of nitrazepam in humans. Seven healthy volunteers received 2 nitrazepam-containing tablet preparations (5 mg) and plasma concentrations were determined regularly from 15 min to 80 h following drug administration. The mean elimination half-life of nitrazepam was 27 h (range 13-34 h). Considerable intra-individual differences in peak level times between the 2 preparations were observed, whereas the extent of bioavailability was rather similar.