Determination of 2, 4-dichlorophenoxyacetic acid in air of workplace by high-performance liquid chromatography
Wen, Y.; Fu, Z.; Xu, J.; Tang, S.; Wang, Q.; Li, H.; Xie, G.; Zhu, Y.; Gu, Y.; Tan, F.
Zhonghua Lao Dong Wei Sheng Zhi Ye Bing Za Zhi 32(6): 458-459
2014
ISSN/ISBN: 1001-9391 PMID: 25169236 Document Number: 677311
To develop a method for determination of 2, 4-dichlorophenoxyacetic acid (2, 4-D) in the air of workplace by high-performance liquid chromatography. 2, 4-D was collected by ultrafine glass filters, desorbed by methanol, separated by a C18 column, and detected by a UV detector. Identification and quantification of 2, 4-D were performed by retention time and peak areas, respectively. The linear range of the test was 2∼200 µg/ml; the elution efficiency was 94.6%- 95.9%; the limit of detection (S/N = 3) was 0.034 µg/ml (injection volume of 20 µl eluant); the lower limit of quantification (S/N = 10) was 0.11 µg/ml; the minimum detectable concentration was 0.011 mg/m(3); the minimum quantifiable concentration was 0.037 mg/m(3) (with sampled air volume of 45 L). This method is convenient and simple in sample collection and preparation, and satisfies all methodological requirements. Therefore, this method is useful for the determination of 2, 4-D in the air of workplace.