Determination of valproic acid by on-line coupled capillary isotachophoresis with capillary zone electrophoresis with conductometric detection

Budáková, L.; Brozmanová, H.; Kvasnicka, F.; Grundmann, M.

Ceska a Slovenska Farmacie Casopis Ceske Farmaceuticke Spolecnosti a Slovenske Farmaceuticke Spolecnosti 56(5): 249-253

2007


ISSN/ISBN: 1210-7816
PMID: 18064807
Document Number: 606368
The paper aimed to test suitable conditions for the determination of VPA in serum using the ITP-CZE method with conductometric detection. It attempted to find whether for the method described by Olvecká et al. it is possible to use a similar procedure of sample preparation as for the gas chromatographic method which is routinely employed for the assessment of VPA in serum. A 200 rl sample of serum was extracted with 200 rl of acetone and after centrifugation 200 rl of the upper layer was withdrawn and supplemented with demineralized water to make 10 ml. When a microsyringe is used for dosing, consumptions of serum can be reduced up to 5 rl, which is advantageous mainly for therapeutic drug monitoring in children. The length of analysis was 20 minutes. The method was linear within the range of 20-400 mg/l, which includes the therapeutic range of VPA (50-100 mg/l). The detection and quantification limits were 2.5 and 8.5 mg/l, respectively. The variation coefficients were below 10% in all three concentrations (30, 70 and 120 mg/l). The new method was compared with the method of gas chromatography by means of the Passing-Bablok regression analysis and both methods were found to give identical results.