Effect of Prompt Inoculator on results of Sceptor system minimal inhibitory concentration determinations for Listeria monocytogenes

Dillon, P.A.; Overman, T.L.

Archives of Pathology and Laboratory Medicine 112(2): 163-165

1988


ISSN/ISBN: 0003-9985
PMID: 3122704
Document Number: 308835
The minimal inhibitory concentrations (MICs) of eight antimicrobial agents were determined for each of 36 isolates of Listeria monocytogenes with the Sceptor system. Two different inoculation procedures, the standard Sceptor log-phase broth (LB) culture and the Prompt Inoculator (PI) (Bauer-Kirby type), were used. The PI MIC/LB MIC ratio (PI/LB ratio) was determined for each antimicrobial agent for each of the isolates. Of the 217 on-scale PI/LB ratio results, all were within the expected range of 0.5 to 2.0. The PI was found to be a convenient, rapid, and suitable method of preparing an inoculum of L. monocytogenes for use with the Sceptor system and should function equally well when testing L. monocytogenes isolates with other commerically available MIC systems.

Document emailed within 1 workday
Secure & encrypted payments