Phantoms for quality assurance in quantitative studies with Scheimpflug photographs

Mayer, H.; Poganatz, J.

Biomedizinische Technik. Biomedical Engineering 32(9): 201-203

1987


ISSN/ISBN: 0013-5585
PMID: 3676431
Document Number: 296241
Quantitative evaluations of Scheimpflug photographs of the anterior eye segment revealed a high scatter of the measured values. So far, it has not been possible to distinguish between methodological errors and errors caused by the patient. This paper presents a model for investigating the methodological error of the device. The evaluation of the Scheimpflug photographs confirmed the assumption: the inconstant illumination of the slit-beam of the Scheimpflug camera is the main source of methodological error.