Multichannel thermistor cryo-osmometer for determination of osmolarity of biological fluids

Kochetygov, N.I.; Gerbut, K.A.; Selivanov, E.A.; Sokolov, I.M.

Patologicheskaia Fiziologiia i Eksperimental'naia Terapiia 1978(2): 81-83

1978


ISSN/ISBN: 0031-2991
PMID: 673487
Document Number: 127387
Schematics and a description are presented for a multi-channel thermistor cryoosmometer which was developed at the Leningrad Research Institute of Hematology and Blood Transfusion [USSR]. The apparatus was used in the determination of animal plasma and urine osmolarity during various traumas and for analysis of plasma substitute osmolarity. Osmolarity determination required 1-2 min and a small amount (0.2 ml) of sample. Osmolarity measurements of samples did not require preliminary calibration relative to standard solutions which eliminated problems in preparation and storage of standardized solutions.