Evaluation of signal noise ratio on analysis of clear cell renal cell carcinoma using DWi with multi-b values

Ding, J.; Xing, W.; Chen, J.; Dai, Y.; Sun, J.; Li, D.

Zhonghua Yi Xue Za Zhi 94(3): 171-173

2014


ISSN/ISBN: 0376-2491
PMID: 24731455
Document Number: 678028
To explore the influence of signal noise ratio (SNR) on analysis of clear cell renal cell carcinoma (CCRCC) using DWI with multi-b values. The images of 17 cases with CCRCC were analyzed, including 17 masses and 9 pure cysts. The signal intensity of the cysts and masses was measured separately on DWI for each b value. The minimal SNR, as the threshold, was recorded when the signal curve manifest as the single exponential line. The SNR of the CCRCC was calculated on DWI for each b value, and compared with the threshold by independent Two-sample t Test. The signal decreased on DWI with increased b factors for both pure cysts and CCRCC. The threshold is 1.29 ± 0.17, and the signal intensity of the cysts on DWI with multi-b values shown as a single exponential line when b ≤ 800 s/mm(2). For the CCRCC, the SNR is similar to the threshold when b = 1 000 s/mm(2) (t = 0.40, P = 0.69), and is lower when b = 1 200 s/mm(2) (t = -2.38, P = 0.03). The SNR should be sufficient for quantitative analysis of DWI, and the maximal b value is 1000 s/mm(2) for CCRCC.