Influence of mimic cardiac rate on hydrodynamics of different mechanical prosthetic cardiac valves in vitro
Chu, Y.-p.; Cheng, J.-l.; Chen, R.-k.; Fan, Y.-b.; Pu, F.
Chinese Medical Sciences Journal 20(2): 138-141
2005
ISSN/ISBN: 1001-9294 PMID: 16075758 Document Number: 591069
To assess the influence of mimic cardiac rate on hydrodynamics of different mechanical prosthetic cardiac valves. US-made CarboMedics bileaflet valve, China-made Jiuling bileaflet valve and C-L tilting disc valve were tested via a pulsatile flow simulator in the aortic position. Testing conditions were set at mimic cardiac rates of 55 bpm, 75 bpm, 100 bpm with a constant mimic cardiac output of 4 L/min. The mean pressure differences (deltaP), leakage volumes (L(E)V) and closing volumes (C(L)V) across each valve, and effective orifice areas (EOA) were analyzed. Within physiological range, deltaP, L(E)V, and C(L)V decreased as mimic cardiac rate increased, with a large extent of variance. EOA increased along with an increase in mimic cardiac rate. It was a different response in terms of cardiac rate alteration for different types of mechanical prosthetic cardiac valves. Mimic cardiac rate change affects hydrodynamics of mechanical prosthetic cardiac valves. Within physiological range, the hydrodynamic of prosthetic bileaflet valve is better than that of tilting disc valve.