Liquid chromatographic analysis of vitamin B6 in soy-based infant formula

Mann, D.L.; Chase, G.W.; Eitenmiller, R.R.

Journal - Association of Official Analytical Chemists 84(5): 1593-1599

2001


ISSN/ISBN: 0004-5756
PMID: 11601481
Document Number: 527143
A liquid chromatographic (LC) method is described for determination of total vitamin B6 in soy-based infant formula. Total vitamin B6 is quantitated by using ion-pair LC after precolumn transformation of phosphorylated and free vitamers into pyridoxol. The limit of detection is 0.3 ng and the limit of quantitation is 1.0 ng on-column (injection volume = 100 microliter). Linear response ranged from 39 to 616 ng/mL (r2 = 0.99986). Analysis of a soy-based infant formula control fortified at 6 different concentration levels gave recoveries that averaged 104%. Assay of SRM 1846 gave results within the certified range (8.6 +/- 0.086 mg/kg versus the certified value of 8.4 +/- 1.0 mg/kg). The method provides a rapid and specific assay for the analysis of total vitamin B6 in fortified soy-based infant formula.