Blood/air partition coefficient determinations: automation and improvement of the equilibrium partitioning in closed systems method

Dills, R.L.; Ackerlund, W.S.; Kalman, D.A.; Morgan, M.S.

Journal of Exposure Analysis and Environmental Epidemiology 3(4): 471-489

1993


ISSN/ISBN: 1053-4245
PMID: 8173346
Document Number: 410347
The equilibrium partitioning in closed systems method has been frequently used for determining partition coefficients between liquid and gas phases. We developed several improvements in this method: preparation of test atmospheres with a dynamic solvent vapor generator, automation of gas phase analysis, and estimation of the precision of the partition coefficients using a statistical procedure that did not require pairing of test and reference measurements. Blood/air partition coefficients measured in a blood-bank blood sample for acetone, 1,1,1-trichloroethane, toluene, and styrene were: 309 +/- 2, 5.6 +/- 0.1, 16.2 +/- 0.5 and 49 +/- 2 (geometric mean +/- estimated SE, respectively). Water/air partition coefficients were: 361 +/- 7, 2.9 +/- 0.3, 4.6 +/- 0.5, and 6.9 +/- 0.9 for acetone, 1,1,1-trichloroethane, toluene, and styrene.

Document emailed within 1 workday
Secure & encrypted payments