Automatic incubator for use with modified A-1 test for enumerating fecal coliform bacteria in shellfish growing waters

Cook, D.W.

Journal - Association of Official Analytical Chemists 64(3): 771-773

1981


ISSN/ISBN: 0004-5756
PMID: 7016826
Document Number: 172472
An automatic water bath incubator which can be constructed from easily obtainable materials is described. This incubator provides temperatures necessary for both the 35 +/- 0.5 degrees C resuscitation period and the 44.5 +/- 0.2 degrees C final incubation period as required in the AOAC method for enumerating fecal coliforms by the modified A-1 test. Statistical analysis of experimental data indicated no significant differences (P greater than 0.50) in fecal coliform counts obtained by using the automatic incubator and the AOAC incubation method.