Calibration and characterization of heat flow transducers for use in hyperbaric helium

Nuckols, M.L.; Piantadosi, C.A.

Undersea Biomedical Research 7(4): 249-256

1980


ISSN/ISBN: 0093-5387
PMID: 7233619
Document Number: 157305
A calibration technique for heat flow transducers (HFT) verified to within 5% by the National Bureau of Standards was used to show significant errors in calibration constants supplied by the manufacturer of the transducers. The effects of variable ambient temperature and hyperbaric helium on HFT signal response were defined. Application of accurate calibration constants and appropriate temperature correction factors when using HFT in hyperbaric helium allowed measurement of convective heat flow [for the assessment of convective human heat loss in cold hyperbaric conditions (diving)] to within 10%.