Percutaneous repair of aortic puncture with Amplatzer closure device during attempted transseptal puncture

Webber, M.R.; Stiles, M.K.; Pasupati, S.

Journal of Invasive Cardiology 25(5): E110-E113

2013


ISSN/ISBN: 1557-2501
PMID: 23645057
Document Number: 670653
Attempted atrial transseptal puncture in a 63-year-old man undergoing an ablative procedure for atrial fibrillation was complicated by inadvertent delivery of an 8 Fr sheath across the aorta. Due to obesity-related perioperative risks, we opted for percutaneous repair rather than open-heart corrective surgery. Our case is unique for the novel percutaneous delivery of an Amplatzer atrial septal defect (ASD) closure device to the defect in the non-coronary aortic cusp through an 8 Fr left atrial multipurpose sheath not designed for this purpose. At 9-month follow-up, he had a mild residual internuclear ophthalmoplegia.

Document emailed within 1 workday
Secure & encrypted payments