GlideScope AVL single-use video laryngoscope
Hirabayashi, Y.
Masui. Japanese Journal of Anesthesiology 62(9): 1139-1142
2013
ISSN/ISBN: 0021-4892 PMID: 24063145 Document Number: 669681
The aim of this case series was to evaluate the performance of GlideScope AVL single-use video laryngoscope for the visualization of the glottis and endotracheal intubations. Visualization of the glottis with GlideScope AVL single-use video laryngoscope was performed in 200 consecutive patients who required tracheal intubation for surgery. The time to complete instrumentation and the visualization of the glottis according to the Cormack-Lehane grade was recorded. GlideScope AVL single-use video laryngoscope provided the grade I view of the glottis in 156 patients and the gradeII view in 44 patients. Intubation with GlideScope AVL single-use video laryngoscope was successful in 199 of the 200 patients. The mean (+/- SD) time for instrumentation was 38 +/- 23 s for orotracheal intubation (n = 170) and 36 +/- 16 s for nasotracheal intubation (n = 28). GlideScope AVL single-use video laryngoscope could be an effective aid for airway management in surgical patients.