Method development and validation for determining 1,3-butadiene in human blood by gas chromatography-mass spectrometry and head-space gas chromatography

Zhang, S.-J.; Shen, B.-H.; Zhuo, X.-Y.

Fa Yi Xue Za Zhi 29(2): 110-115

2013


ISSN/ISBN: 1004-5619
PMID: 23930504
Document Number: 669542
To develop a simple, validated method for identifying and quantifying 1,3-butadiene (BD) in human blood by gas chromatography-mass spectrometry (GC-MS) and head-space gas chromatography (HS-GC). BD was identified by GC-MS and HS-GC, and quantified by HS-GC. The method showed that BD had a good linearity from 50 to 500 microg/mL (r > 0.99). The limits of detection and quantification were 10 microg/mL and 50 microg/mL, respectively. Both the intra-day precision and inter-day precision were < 6.08%, and the accuracy was 96.98%-103.81%. The method was applied to an actual case, and the concentration of BD in the case was 242 microg/mL in human blood. This simple method is found to be useful for the routine forensic analysis of acute exposure to BD.