Migrated Essure permanent birth control device: sonographic findings
Khati, N.J.; Gorodenker, J.; Brindle, K.A.
Journal of Clinical Ultrasound Jcu 42(4): 223-226
2014
ISSN/ISBN: 1097-0096 PMID: 24122949 DOI: 10.1002/jcu.22096Document Number: 624564
We report a case of a migrated Essure permanent birth control device. The correct diagnosis was made on conventional two-dimensional and three-dimensional pelvic sonography 7 years after placement of the device when the patient presented with persistent right-sided pain. The 3-month post placement hysterosalpingogram had shown an appropriately occluded right fallopian tube but had overlooked the abnormal position of the right Essure device, which was too proximal and extending slightly in the uterine cavity.