Accurate early infant HIV diagnosis in primary health clinics using a point-of-care nucleic acid test

Jani, I.V.; Meggi, B.; Mabunda, N.éd.; Vubil, A.; Sitoe, N.E.; Tobaiwa, O.; Quevedo, J.I.; Lehe, J.D.; Loquiha, O.; Vojnov, L.; Peter, T.F.

Journal of Acquired Immune Deficiency Syndromes 67(1): E1-E4

2014


ISSN/ISBN: 1944-7884
PMID: 24933096
DOI: 10.1097/qai.0000000000000250
Document Number: 622390
To evaluate the accuracy of a point-of-care (POC) nucleic acid-based test (NAT) for early infant HIV diagnosis (EID) in primary health clinics in Mozambique. POC and laboratory NAT EID tests were conducted on matched blood samples collected from 827 HIV-exposed infants younger than 18 months who were enrolled consecutively at 4 periurban primary health clinics and the central hospital in Maputo. Lancet heel draw blood collected by nurses was tested on site for HIV-1/-2 RNA on the Alere HIV NAT POC device and also used to create dried blood spots for later laboratory EID testing on the Roche Cobas Taqman/Ampliprep instrument. Results were used to determine the sensitivity, specificity, and agreement between the POC and laboratory NAT EID tests. The sensitivity and specificity of POC NAT EID testing were 98·5% (95% confidence interval (CI): 91.7 to 99.9, n = 65) and 99·9% (95% CI: 99.3 to 100, n = 762), respectively, compared with laboratory EID tests. Overall agreement was high (Cohen kappa = 0·981; 95% CI: 0.96 to 1.00). Positive (98·5%; 95% CI: 96·3 to 100) and negative 99.9% (95% CI: 99.7 to 100) test agreement was also high. Primary health care nurses accurately performed POC NAT EID testing within primary health care clinics. On-site nucleic acid-based EID testing is technically feasible in clinic settings and could be used in efforts to improve access to pediatric HIV antiretroviral treatment.

Document emailed within 0-6 h
Secure & encrypted payments