Automatic measurement of skin wheals provoked by skin prick tests

Prinz, M.; Vigl, K.; Wöhrl, S.

Studies in Health Technology and Informatics 116: 441-446

2005


ISSN/ISBN: 1879-8365
PMID: 16160297
Document Number: 588931
Skin prick tests (SPT) represent the standard method for the diagnosis of type-1 allergies. The skin wheals provoked by the SPTs are considered positive above certain cut-off diameters, usually 3 mm. At present their size is mostly estimated by measuring the diameter with rulers. Since the shape of wheals usually is not circular, the measurement of diameters leads to imprecise results. Therefore, we developed an algorithm for precisely measuring the wheals' area in mm(2) automatically. The average deviation of the automatic measurement on a test set achieved with the developed algorithm was 6.9 %. Compared to the maximum standard deviation of 6.59 % when measuring the manually redrawn outlines, the automatic method works sufficiently well.