Head-neck disassembly of an uncemented revision stem treated by addition of a proximal spacer

Regis, D.; Modena, M.; Danesi, G.; Corallo, F.; Bartolozzi, P.

Acta Orthopaedica Belgica 69(5): 463-466

2003


ISSN/ISBN: 0001-6462
PMID: 14648959
Document Number: 566596
A 67-year-old woman underwent revision total hip replacement with a long, distally fixed uncemented stem inserted through a transfemoral approach. Initial stability was not achieved and the stem subsided early. Probably due to a conflict with the minor trochanter, the prosthetic head detached from its neck. Instead of a re-revision of the stem, a proximal spacer was inserted to restore stability and to compensate for leg-length discrepancy. This has proved to be a stable and satisfactory solution after three years follow-up.