Use of intraoperative fluoroscopy during laparotomy to identify fragments of retained Essure microinserts: case report

Howard, D.L.; Christenson, P.J.; Strickland, J.L.

Journal of Minimally Invasive Gynecology 19(5): 667-670

2012


ISSN/ISBN: 1553-4650
PMID: 22935312
DOI: 10.1016/j.jmig.2012.04.007
Document Number: 564376
In previous case-reports of Essure microinsert perforation, the microinsert was successfully removed at laparoscopy. Herein is discussed the scenario of persistent pelvic pain over several years after an apparently successful laparoscopic retrieval of a perforating right-sided microinsert. In the interim, the patient underwent 2 unsuccessful exploratory laparotomy procedures in an attempt to retrieve additional microinsert fragments that had perforated the uterus. Successful management of Essure microinsert perforation in this patient ultimately required use of intraoperative fluoroscopy. Surgeons performing laparoscopy or laparotomy to retrieve Essure microinserts that have perforated should be aware that these are not always visible to the naked eye, and there should be a low threshold to use intraoperative fluoroscopy to ensure that all perforating fragments have been removed.