Correlation between Goldmann and non-contact tonometry based on corneal thickness

Schepens, G.; Urier, N.; Bechetoille, A.; De Potter, P.

Bulletin de la Societe Belge d'Ophtalmologie 279: 15-19

2001


ISSN/ISBN: 0081-0746
PMID: 11344711
Document Number: 533530
In this transversal study, we measure the intra ocular pressure by means of the Goldmann tonometer and a Non-Contact tonometer along with the central cornea thickness in 136 eyes of 73 patients. The statistical analysis of the collected data doesn't allow us to establish a correlation between the variation of the corneal thickness and the difference between the Goldmann tonometer and Non-Contact tonometer measures. The relative precision of the Non-Contact tonometer compared with the Goldmann tonometer doesn't seem influenced by the central cornea thickness.