LASIK--comparison of oscillating and rotating microkeratotomy using scanning electron microscopy

Stodůlka, P.; Vaculík, J.; Rozsíval, P.; Lichnovský, V.

Ceska a Slovenska Oftalmologie Casopis Ceske Oftalmologicke Spolecnosti a Slovenske Oftalmologicke Spolecnosti 55(5): 304-311

1999


ISSN/ISBN: 1211-9059
PMID: 10596481
Document Number: 502915
Scanning electron microscopy was used to evaluate the differences in the quality of the cut produced by oscilating microkeratome Flapmaker and Draeger rotating microkeratome. Higher quality of the edge of the cut was found with Flapmaker. Both microkeratomes produce uniform surface of the cut. Chatter is much less expressed with Flapmaker. No difference at the interface tissue was found with SEM three weeks postoperatively comparing the microkeratomes.