The design and clinical application of an apparatus for monitoring neuromuscular transmission function during operation under general anaesthesia

Lin, S.; Zhou, W.; Yang, J.; Li, Y.; Wang, Y.; Wang, Q.

Sheng Wu Yi Xue Gong Cheng Xue Za Zhi 15(2): 123-127

1998


ISSN/ISBN: 1001-5515
PMID: 12548897
Document Number: 493558
This paper introduces a monitoring system of musle relaxant in action during the operation under general anaesthesia. The design of the monitor is based on measurement of acceleration. It consists of an acceleration transducer, a nerve stimulator and a microcomputer unit. The transducer can be fastened to the thumb, and the signal of the thumb movement in response to ulnar nerve stimulation is measured and registered in the computing unit. In clinical practice, it has been proved that this new apparatus can achieve reliable evaluation of nuromuscular transmission function and improve the quality of anesthesia in patients care during surgical operation.