Determination of platinum in wine by graphite furnace atomic absorption spectrometry

Cabrera-Vique, C.; Teissedre, P.L.; Cabanis, M.T.; Cabanis, J.C.

Journal - Association of Official Analytical Chemists 80(1): 57-62

1997


ISSN/ISBN: 0004-5756
PMID: 9011059
Document Number: 483775
A method based on graphite furnace atomic absorption spectrometry (GFAAS) was developed for determining platinum in wine. Wine samples were prepared by microwave acid digestion or dry mineralization. The method of standard addition was used for Pt determination in untreated wine samples and mineralized samples. Analyte modifiers and furnace conditions were optimized. Effects of cations (Mg2+, Ca2+, K+, Na+, and NH4+) and anions (PO4(3)-, SO4(2)-) were tested separately and in combination. Analytical characteristics of the method were optimized for analyte recovery and signal enhancement. Recoveries ranged from 92.5 to 102%, and precision reproducibility relative standard deviation varied from 7.5 to 10%. Red, rosé, and white wines from France were analyzed. Platinum levels found in most wines were very low (< 10 micrograms/L).