A simple, lightweight device for measurement of hemoglobin; the HemoCue Blood Hemoglobin Test
Nishiyama, T.; Hanaoka, K.
Masui. Japanese Journal of Anesthesiology 46(2): 284-285
1997
ISSN/ISBN: 0021-4892 PMID: 9071118 Document Number: 477500
In some situations such as a sudden bleeding during surgery, rapid testing of blood hemoglobin concentration is necessary. The HemoCue Blood Hemoglobin Test uses a lightweight photometer, which is easily movable anywhere. The HemoCue needs only 10 microliters of blood and it takes only about 45 seconds to produce the result. Therefore, the device might be useful in the operating suite and the emergency unit. To evaluate the accuracy of the HemoCue, we compared the blood hemoglobin concentrations measured by the HemoCue with those measured by the ABL 300. A positive correlation was found between the variables by the HemoCue (Y) and those by the ABL 300 (X); Y = 0.944X - 0.208, r = 0.97, P < 0.001. It is concluded that the HemoCue is a reliable device for measurement of blood hemoglobin concentration.