Enzyme immunoassay for erythrocyte aldose reductase

Nishimura, C.; Hamada, Y.; Tachikawa, T.; Ishikawa, T.; Gui, T.; Tsubouchi, J.; Hotta, N.; Tanimoto, T.; Urakami, T.

Clinical Chemistry 40(6): 889-894

1994


ISSN/ISBN: 0009-9147
PMID: 8087983
Document Number: 431340
This two-site immunoassay measures erythrocyte aldose reductase by using monoclonal and polyclonal antibodies to recombinant human enzyme. Total incubation time is 2.5 h, and the limit of detection is < 0.05 microgram/L. Analytical recovery tested with blood samples from healthy and diabetic individuals was 101-106%. Average CVs within and between assays were 3.7% and 4.8%, respectively. The enzyme content determined by this system correlated well with the activity of aldose reductase isolated from the same erythrocyte preparations. The amount of erythrocyte aldose reductase per milligram of hemoglobin was higher in women than in men (P < 0.001), but no significant correlation was observed between the amount of enzyme and the age of the individuals. This assay method should provide useful clinical information to optimize administration of aldose reductase inhibitors for effective prevention and treatment of diabetic complications.