Reduction of NMR image artefacts by using optimal materials as diagnostic aids and implants

Fritzsche, S.; Thull, R.; Haase, A.

Biomedizinische Technik. Biomedical Engineering 39(3): 42-46

1994


ISSN/ISBN: 0013-5585
PMID: 8193244
Document Number: 426617
Differences in magnetic susceptibility, as occur for example at the boundary between implant materials and the surrounding body tissue, result in artefacts and a signal loss in magnetic resonance imaging. By using materials with a magnetic susceptibility matched to the respective environment, it is possible to minimize both artefacts and signal loss. Such materials can be made by combining two materials of different magnetic susceptibility (e.g. diamagnetic and paramagnetic materials) in such a way that the resulting material has the desired effective magnetic susceptibility.