Evaluation of a hyperbaric system to be used in conjunction with a fluorometer
Colton, J.S.; Grossman, Y.; Miller, K.; Pocotte, S.L.
Undersea and Hyperbaric Medicine Journal of the Undersea and Hyperbaric Medical Society Inc 20(4): 375-382
1993
ISSN/ISBN: 1066-2936 PMID: 8286990 Document Number: 411870
A high-pressure chamber that can be used inside the sample chamber of a spectrofluorometer is described and some performance characteristics are presented. The chamber body, constructed of 316 stainless steel, is temperature regulated using resistive heating elements and a microprocessor-based proportional integral derivative controller. The chamber holds a standard 1-cm2 cuvette that indexes with an electromagnetic stirrer. Injection of different solutions into the closed and pressurized (6.8 MPa) vessel is accomplished by computer-controlled, low-volume solenoids attached to separate microliter injection ports. Repetitive injections of fluids down to a volume of 7 microliters are possible in the pressurized chamber. Temperature stability of the chamber is +/- 0.2 degrees C at atmospheric or elevated pressure. However, during the initial phase (first 3 min) of pressurization, at a compression rate of 0.62 MPa/min, a 0.23 degrees C/min increase in temperature occurs. The chamber windows depress the relative intensity of the emitted light by approximately 20% for visible light and 40% for near UV; however, total sensitivity of the system is sufficient to accomplish most determinations while maintaining a good signal-to-noise ratio. This system can be used to evaluate the response of several molecular and cellular events during compression and at depth with the use of various fluorometric probes.