A simple radioassay for detection of antithyroglobulin autoantibodies

Kumar, A.; Shah, D.H.

Indian Journal of Medical Research 96: 344-349

1992


ISSN/ISBN: 0019-5340
PMID: 1283857
Document Number: 391468
Radioassay technique for detection of anti-thyroglobulin autoantibodies (ATA) in serum samples using protein-A rich Staphylococcus aureus as an immune complex separating agent was comparable (r = 0.99; n = 42; P < 0.001) to that evaluated with anti-human gammaglobulin (AHGG). The inter- and intra-assay coefficient of variations were lower (7.4 and 2.2% respectively) using Staph. aureus as compared to that observed with AHGG (10.4 and 4.7% respectively). A highly significant correlation was observed (r = 0.53; n = 71; P < 0.001) between per cent bound radiolabel thyroglobulin in radioassay and log reciprocal titre of ATA by haemagglutination method. Radioassay scored more number of positive sera as compared to haemagglutination method for normal controls (6 vs 1; n = 46) and patients of thyroid diseases (151 vs 70; n = 238). Due to the relatively poor stability of tracer on storage larger inter-assay coefficient of variations were observed with the radiolabelled preparations older than 10-12 days. The radioassay method evaluated is sensitive, reproducible and useful in detecting presence of anti-thyroglobulin autoantibodies.