A catheter-based enzyme-coupled electrode for measurement of lactate

Waite, R.I.; Waite, L.R.; Lim, S.P.; Spieker, C.

Biomedical Instrumentation and Technology 25(6): 461-464

1991


ISSN/ISBN: 0899-8205
PMID: 1777769
Document Number: 368129
An enzyme-coupled electrode was incorporated into a catheter to measure the concentration of lactate solutions. The sensor output was found to be linearly dependent on the lactate concentration. The average nonlinearity over a sample of nine catheters was 5.88% full scale. Hysteresis for the device was low, with an average of 4.54% full scale. Preliminary results indicate that in-vivo measurements may be possible, which are not possible with other lactate measurement methods.