A comparison of three different techniques for measuring glomerular filtration rate

Herbst, C.P.; Dunn, M.; Viviers, A.C.; van Zyl, M.; van Aswegen, A.; Otto, A.C.

South African Medical Journal 78(5): 266-267

1990


ISSN/ISBN: 0256-9574
PMID: 2118278
Document Number: 350317
The clearance of inulin, creatinine and radioactive tracers from the blood may be used to measure glomerular filtration rate (GFR). These techniques, however, are usually invasive and time-consuming. Although the clearance of a radioactive tracer is usually applied in nuclear medicine for the determination of GFR, it is also possible to convert the concentration of the tracer in the kidneys to absolute GFR by means of a regression equation. As this new technique is much faster, we have compared it with the conventional technique. A good correlation was found with the standard radionuclide techniques (r = 0.91), but the reference method was under-estimated on the average by 14 ml/min. The new regression equation derived in our clinic will ensure future accurate GFR measurements within 6 minutes.