Difficulties during removal of fluted femoral intramedullary rods
Star, A.M.; Whittaker, R.P.; Shuster, H.M.; Duda, J.; Menkowitz, E.
Journal of Bone and Joint Surgery. American Volume 71(3): 341-344
1989
ISSN/ISBN: 0021-9355 PMID: 2925705 Document Number: 332805
The difficulties that were encountered during removal of a fluted intramedullary femoral rod from six patients were reviewed. The mechanisms of failure included metal breakage at three different locations along the rod or the rod extractor. Three of the six rods were left in place due to difficulties in removal. The design of the fluted intramedullary rod appears to be the main cause of the problem.