Thyrotroph function assessed by sensitive measurement of thyrotropin with three immunoradiometric assay kits: analytical evaluation and comparison with the thyroliberin stimulation test
Wians, F.H.; Jacobson, J.M.; Dev, J.; Heald, J.I.; Ortiz, G.
Clinical Chemistry 34(3): 568-575
1988
ISSN/ISBN: 0009-9147 PMID: 3127084 Document Number: 321624
We evaluated the analytical and clinical performance of three "sensitive" immunoradiometric assay (IRMA) kits (Tandem-R TSH HS, Hybritech, Inc.; EchoClonal TSH, Bio-Rad; Coat-A-Count TSH IRMA, Diagnostic Products Corp.) for measurement of thyrotropin (TSH) and compared their performance against a "regular" IRMA (ARIA-HT, Becton-Dickinson) to determine whether these assays might eliminate the need to perform the thyroliberin (TRF) stimulation test. We concluded that the Tandem and EchoClonal kits may obviate the need to perform TRF stimulation tests in some patients. Using only the basal TSH concentration to predict the TSH response to TRF, we found all three sensitive TSH assays to be useful for detecting abnormal thyrotroph function. Dose-response, linearity, analytical recovery, and specificity were acceptable for all kits, but intra-assay precision at very low TSH concentrations and analytical sensitivity differed considerably among the kits. Using the EchoClonal assay, we established a normal reference interval for TSH of 0.4-4.6 milli-int. units/L.