Determination of sulfite in foods by headspace liquid chromatography
Lawrence, J.F.; Chadha, R.K.
Journal - Association of Official Analytical Chemists 71(5): 930-933
1988
ISSN/ISBN: 0004-5756 PMID: 3235411 Document Number: 310801
Sulfite was determined in a variety of foods by liquid chromatography (LC) after the samples were mixed with a solution containing mannitol, FeSO4, and Na2HPO4, adjusted to pH 11, and left to stand for 15 min at room temperature. An aliquot of the mixture was placed in a headspace vial and mixed with 50% H3PO4. After 15 min, a portion of the headspace was removed with a syringe containing LC mobile phase without acetonitrile. The syringe was shaken and an aliquot of the solution was analyzed on an anion exchange column with a mobile phase of 0.03M methane sulfonate (pH 10.8) containing 5% acetonitrile. Sulfite was detected amperometrically (glassy carbon electrode) at +0.7 V. The method was successfully compared to the FDA-modified Monier-Williams procedure for a variety of foods. Minimum detectable levels were about 1 microgram/g, based on a 15 g sample.