Analysis of the causes of potential wear and breakdown of elements of artificial heart valves of the disk type

Dobrova, N.B.; Iofis, N.A.; Kozyrkin, B.I.; Agafonov, A.V.; Zaretskiĭ, I.V.

Meditsinskaia Tekhnika 2: 8-12

1987


ISSN/ISBN: 0025-8075
PMID: 3586967
Document Number: 300523
The reliability analysis of the artificial cardiac valve and experiments made it possible to establish that the ACV reliability is ensured by the physical and mechanical properties of the material, technology observance and thorough control over each operation during the manufacturing process. The accelerated testing of the new MX disk valve made in the USSR showed a satisfactory degree of the disk wear and the absence of breaks-down after 400 million working cycles which is equivalent to ten-years' work in a human body.

Document emailed within 1 workday