Image quality parameters of film-screen systems

Hoeschen, D.

Rontgen-Blatter; Zeitschrift für Rontgen-Technik und Medizinisch-Wissenschaftliche Photographie 40(6): 193-199

1987


ISSN/ISBN: 0300-8592
PMID: 3616429
Document Number: 295645
The image quality in radiology is to a large extent determined by three measurable parameters of screen-film systems: the H-D curve, the modulation transfer function and the Wiener spectrum. Measuring methods of the three parameters are described and results are presented. It turns out that the modulation transfer function is correlated to the visual sharpness and the Wiener spectrum is correlated to the visual impression of graininess. The combination of the three parameters to a single image quality criterion in the form of the detective quantum efficiency (DQE) demonstrates that the DQE is not a generally valid quality criterion for high sharpness and low graininess.