A gamma-ray sensitometer for film-screen combinations in diagnostic radiology

Nowotny, R.; Bliem, J.

Rontgen-Blatter; Zeitschrift für Rontgen-Technik und Medizinisch-Wissenschaftliche Photographie 38(4): 117-120

1985


ISSN/ISBN: 0300-8592
PMID: 4001775
Document Number: 242861
On account of the high cost of standardised x-ray equipment, a 5 Ci Am-241 gamma ray source (59.6 keV) was used for setting up a sensitometer. The exposure rate at the film cassette is adjusted by means of a variable distance (5-250 cm) from the source. Using exposure times of 2 sec, screens with sensitivities greater than 500 can be measured. The properties of the sensitometer are described.