Automation of microtiter plate-chromogenic substrate LAL endotoxin assay method by use of a modified Cetus Pro/Pette Express system

Martin, P.A.; Gaunnac, G.L.; Tsuji, K.

Journal of Parenteral Science and Technology a Publication of the Parenteral Drug Association 40(2): 61-66

1986


ISSN/ISBN: 0279-7976
PMID: 3723252
Document Number: 283099
An automated microtiter plate-chromogenic substrate LAL system for the assay of bacterial endotoxin has been developed using a Cetus Pro/Pette EXPRESS system, modified with the addition of a solid-state incubator block. The automated system assays 96 samples in approximately 22 min. Sensitivity of the automated system is approximately 0.01 EU/ml with a linear assay range from 0.03 to 0.14 EU/ml and the relative standard deviation (RSD) of approximately 6%. The automated microtiter plate-chromogenic substrate LAL system has been validated for testing a variety of pharmaceutical injectable products and materials. The automated system has been shown to represent significant improvements in speed, accuracy, and precision of the LAL assay method for endotoxin.

Document emailed within 1 workday