Validity and reliability of cardiac output by arterial thermodilution and arterial pulse contour analysis compared with pulmonary artery thermodilution in intensive care unit

Werawatganon, T.; Punyatavorn, S.; Chatkaew, P.; Bunburaphong, P.

Journal of the Medical Association of Thailand 86(Suppl 2): S323-S330

2003


ISSN/ISBN: 0125-2208
PMID: 12930006
Document Number: 2591
Cardiac output measurement has a significat role in the critical care setting. The standard of measurement currently is via pulmonary arterial catheter but it has some technical difficulties and serious complications. The authors performed a new method of measurement that used a catheter in a femoral artery. The results of both methods performed simultaneously in 10 surgical intensive care patients every 2 hours for 24 hours were compared. There was high correlation between the two methods, r = 0.97. The average difference of the cardiac output values was 0.46 l/min with standard deviation 0.56 l/min.