Reliability of Registration in Imageless Navigation Total Hip Arthroplasty in Semilateral Decubitus Position

Suksathien, Y.; Suarjui, J.; Chongmuenwai, A.

Journal of the Medical Association of Thailand 100(4): 457-462

2017


ISSN/ISBN: 1551-8892
PMID: 29911843
Document Number: 17349
One of the limitations in imageless navigation is the variability in bony landmark registration among surgeons. To determine intra and inter-observer reliability of the bony landmark registration in three surgeons with varying The authors reviewed 60 cases receiving cementless THA with imageless navigation between June 2014 and April 2015. All cases were registered for anterior pelvic plane (APP) by three surgeons. The first surgeon (S1-YS) is senior staff with experience in imageless navigation, the second surgeon (S2-JS) is junior staff with one year of experience in imageless navigation (with more than 50 cases), and the third surgeon (S3-AC) is a fourth-year orthopedic resident with no experience in imageless navigation. Using the final cup position of the experienced surgeon as the gold standard to determine the reliability of registration. There were no significant differences in cup abduction angle between (S1) and (S2) (p = 0.27) and (S1) and (S3) (p = 0.79). There was no significant difference in cup anteversion angle between (S1) and (S2) (p = 0.1) but there was significant difference between (S1) and (S3) (p<0.001). For the intra-observer reliability, the ICCs of abduction angle was 0.95 for (S2) and 0.86 for (S3) and the ICCs of anteversion angle was 0.91 for (S2) and 0.86 for (S3). For the inter-observer reliability, the ICCs of abduction angle between (S1) and (S2) was 0.89 and between (S1) and (S3) was 0.87, the ICCs of anteversion angle between (S1) and (S2) was 0.8 and between (S1) and (S3) was 0.72. The reliability of registration was acceptable in abduction angle among surgeons. The reliability of registration

Document emailed within 1 workday
Secure & encrypted payments

Reliability of Registration in Imageless Navigation Total Hip Arthroplasty in Semilateral Decubitus Position