Sterility verification of innovative pressurized carbon dioxide lavage for bone surface preparation in cemented knee arthroplasty

Boontanapibul, K.; Ruangsomboon, P.; Charoencholvanich, K.

Journal of the Medical Association of Thailand 97 Suppl. 9: S108-S111

2014


ISSN/ISBN: 0125-2208
PMID: 25365901
Document Number: 15017
To verify sterility of innovative pressurized carbon dioxide (CO) lavage for bone surface preparation in cemented knee arthroplasty. Twenty-five 500 L. air samples from the innovative pressurized CO, lavage were laboratory tested to verify the presence or absence of either bacteria or fungi in microbial air samplings taken in a biological safety cabinet. CO2 gas was delivered though a pressure regulator, a sterile delivery tube, a hand piece, and a microbial air filter with a pressure level of 50 psi, according to the medical standard. Contamination samples for bacteria and fungi were taken by using Tryptic Soy Agar and SabouraudAgar as the culture media and were then analyzed at 48 hours and 10 days, respectively. Twenty-five samples of microbial culture showed no bacterial or fungal growth on either Tryptic Soy Agar or Sabouraud Agar The sterility of the innovative pressurized carbon dioxide lavage was confirmed by laboratory test results and was approved for application for bone surface preparation in cemented total knee arthroplasty.

Document emailed within 1 workday