Unfocused effect on the measurement of ocular aberrations
Srivannaboon, S.; Adulbhan, N.
Journal of the Medical Association of Thailand 88 Suppl. 9: S31-S36
2005
ISSN/ISBN: 0125-2208 PMID: 16681049 Document Number: 10335
To evaluate the effect of unfocused measurement in the value of ocular aberrations. The present study was conducted at Siriraj Hospital, Mahidol University. Ocular aberrations of 20 eyes in 10 myopic patients (myopia less than -6.00 diopters (D) and astigmatism less than -2.00 diopters (D)) were analyzed by using Zywave aberrometer (Bausch and Lomb, USA). The measurements were done at the pupillary plane (focus) and 6 other different planes (unfocus); 5, 10 and 15 mm inside and outside pupillary plane respectively. The value of each measurement was analyzed to find the effect of unfocused measurement on ocular aberrations. The magnitude of error (root mean square, RMS) of the spherical component of refraction at 5, 10, 15 mm inside and outside the pupillary plane was 0.16, 0.44, 0.57 D and 0.21, 0.38, 1.35D respectively. The RMS of astigmatic component of refraction inside and outside the pupillary plane at the same distance was 0.19, 0.50, 0.80 and 0.18, 0.52, 1.55 D respectively. The RMS of higher order aberrations inside and outside the pupillary plane at the same distance was 0.05, 0.13, 0.15 and 0.06, 0.1, 0.1 microns respectively. There was more effect of outward unfocused measurement in the value of refraction (spherical and astigmatism) than inward unfocused measurement. Higher order aberration showed a similar error in both inward and outward defocus. The magnitude of error was 10 - 12% within 5 mm unfocused planes.