Dislocation of cochlear implant magnet as a complication following MRi

Murtojärvi, S.; Salonen, J.

Duodecim; Laaketieteellinen Aikakauskirja 133(5): 497-500

2017


ISSN/ISBN: 0012-7183
PMID: 29206000
Document Number: 694530
According to current best knowledge, an MRI scan can be performed for patients with cochlear implants. The warnings and recommendations of the implant manufacturers must be followed strictly to prevent complications, such as overheating, migration or demagnetization of the magnet in the implant. We report on a case of cochlear implant magnet dislocation as a complication for an MRI scan. The patient had a tight bandage around the head to hold the magnet in place as recommended by the manufacturer, but apparently the bandage was not in the correct place.