Marked Aneurysmal Expansion Caused by Type II Endoleaks Eight Years after Thoracic Endovascular Aortic Repair;Report of a Case

Midorikawa, H.; Takinami, G.; Kageyama, R.; Ueno, K.; Kanno, M.; Watanabe, M.

Kyobu Geka. Japanese Journal of Thoracic Surgery 70(7): 540-543

2017


ISSN/ISBN: 0021-5252
PMID: 28698425
Document Number: 692058
We report a case of marked aneurysmal expansion by type II endoleaks 8 years after thoracic endovascular aortic repair. The preoperative chest computed tomography (CT) demonstrated the descending thoracic aorta of 95 mm with type II endoleak. We performed suture closure of 3 endoleaks and partial aneurysmorrhaphy via left thoracotomy under partial cardiopulmonary bypass. The postoperative CT showed no endoleak with shrinkage of the aneurysm. The patient was discharged on the 21th postoperative day uneventfully.