Coolant effectiveness in dental cutting with air-turbine handpieces
Leung, B.T.W.; Dyson, J.E.; Darvell, B.W.
New Zealand Dental Journal 108(1): 25-29
2012
ISSN/ISBN: 0028-8047 PMID: 22439318 Document Number: 658036
To establish a strategy for evaluating coolant effectiveness and to compare typical cooling conditions used in dental cutting. A test system comprising a resistive heat source and an array of four type K thermocouples was used to compare the cooling effectiveness of air alone, water stream alone, and an air-water spray, as delivered by representative air-turbine handpieces. Mean temperature change at the four sites was recorded for a range of water flow rates in the range 10 to 90 mL min(-1), with and without air, and with and without the turbine running. The thermal resistance of the system, R, was calculated as the temperature change per watt (KW(-1)). For wet cooling (water stream and air-water spray), R was 5.1 to 11.5 KW(-1), whereas for air coolant alone the range was 18.5 to 30.7 KW(-1). R for air-water spray was lower than for water stream cooling at the same flow rate. The thermal resistivity approach is a viable means of comparative testing of cooling efficacy in simulated dental cutting. It may provide a reliable means of testing handpiece nozzle design, thus enabling the development of more efficient cooling.