Hand-held water fluoride analysis: An accessible caries prevention tool for dental professionals

Quock, R.L.; Yank, S.W.; Chan, J.T.

General Dentistry 59(1): 59-63

2011


ISSN/ISBN: 0363-6771
PMID: 21613041
Document Number: 652523
This study sought to compare the relative accuracy of a commercially available hand-held water fluoride analysis unit with a standard laboratory bench-top fluoride-specific electrode/millivoltmeter apparatus, with the goal of identifying possible practical applications of the hand-held unit for preventive dentistry. The units analyzed identical gravimetrically prepared fluoride solutions ranging from 0.1 to 4.0 ppm. The average difference between the measurements from the hand-held unit and the nominal values of the fluoride solutions was 0.011 ppm (SD = 0.068), and the average difference between the hand-held unit's measurements and the bench-top unit's measurements was 0.030 ppm (SD = 0.115). T-test analysis demonstrated no statistical difference between measurements from the hand-held unit with either the nominal values of the fluoride solutions or the bench-top unit's measurements. Results indicate that the hand-held water fluoride analysis unit has an appropriate level of accuracy for the measurement of fluoride levels in drinking water samples by dental professionals.